GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
3-6
Section 3-3 - Receiving and Unpacking the Equipment
CAUTION:
When using sharp tools to open packing materials, take care to avoid cutting or damaging
the contents.
Note:
Unless otherwise specified, referenced items are shown in
1) Cut and remove the three securing steel strips (item 16).
2) Release the eight Clip-lok
TM
clips (item 15) securing the front wall (item 13) and remove the front
wall. (Refer to the procedure described in
3) Release the eight Clip-lok
TM
clips securing the back wall (item 14) and remove the back wall.
4) Remove the small cartons containing the probes and peripheral options (items D and E in
below) from the wooden crate.
5) Remove the console control support back foam (item 11 in
6) Remove the keyboard support foam located in front of the keyboard (item 10 in
7) Remove the back support foam located in front of the rear wheels (item 9 in
8) Remove the top cover by opening the six clips fastening it to the
right
and
left
walls. For details on
9) Remove the
right
wall (item 6) and the
left
wall (item 7) by opening the three clips fastening each to
the base.
10) Cut and remove the antistatic sheet (item 2 in
below) that is wrapped around the unit,
taking care not to damage the antistatic cover.
11) Remove the antistatic cover (item 4 in
).
Figure 3-3 Probes and Peripherals in Original Packaging