GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
4-10
Section 4-3 - Functional Check
4-3-4-5
Plug and Scan Test
Table 4-17 Plug and Scan Test
Task
Expected Result(s)
1
Press the
On/Off
button on the control console for no
more than 3 seconds.
The system enters
Standby
mode.
2
Turn OFF the system’s circuit breaker.
The system remains in
Standby
mode, indicated by the
blinking
On/Off
button LED.
3
Turn ON the system’s circuit breaker.
Press the
On/Off
Button.
The system returns quickly to the
Power-ON
state -less
than 30 seconds.
4
Turn OFF the system’s circuit breaker. Turn ON the
system’s circuit breaker after entering to Standby mode.
The system enters
Standby
mode and stays in this
mode. The
On/Off
button LED is blinking.
5
Turn OFF the system’s circuit breaker.
Press the
On/Off
button.
The system enters the
Power-OFF
state. The
On/Off
button LED is OFF.