GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics/Troubleshooting
7-11
7-3-2
Radio Frequency Interface (RFI) Diagnostic Tests (for RFI Configuration)
1) Access the
FE Diagnostics
dialog box, as described in the
Accessing the Front End Diagnostic
2) In the
FE Diagnostics
dialog box, trackball to the
RFI
tab and press
Select
,
The
RFI
tab is displayed, as shown below:
Figure 7-5 RFI Tab - FE Diagnostics Dialog Box
Note:
Proper operation of the RFI board may be tested by performing either of the two tests:
Self Test
or
Full Test
, as required. (While the Full Test is in progress, it may be necessary
to wait a while since this requires longer than the Self Test for completion).
3) Verify that the
Comment
area indicates that the system is ready to perform the test (as shown in
4) Trackball to the
Full Test
tab and press
Select
.
OR
Trackball to the
Self Test
tab and press
Select
.
The
Full Test
(or
Self Test
) LED changes color as follows:
•
Yellow:
The test is in progress.
•
Green:
The test has been completed and no problems were identified.
•
Red:
Problems were identified during the performance of the test.
Data Flow Map
Test Description Area
Test Area
Comment Area
Tabs