Chapter 8 - Replacement Procedures
8-27
8) Lift out the keyboard, completely removing it from the machine chassis.
8-3-3-4
Keyboard Installation Procedure
1) Return the keyboard to its original position.
2) Push the keyboard as far in as it can go, carefully aligning the forked metal aligning plate with the
central nut, and then fasten it to the metal frame from underneath at its far end corners with the two
M4 nuts, one on either corner - refer to
3) Return and fasten the two M4 screws to secure the keyboard matrix to the metal frame (one on each
of the
left
and
right
4) Re-connect in the four keyboard cables, located at the bottom of the keyboard - see
5) Return the control console upper cover (front), as described in the
(Front) Installation Procedure
6) Return the control console lower cover, as described in the
7) Return the control console lower rear cover, as described in the