Chapter 8 - Replacement Procedures
8-49
Note:
When replacing a BEP, one must re-install software using the software provided with the
system.
Do NOT use the software supplied with the BEP
unless it is a version later
than the software currently installed on the system.
5) Place the BEP on a table and start removing the required parts (listed below), as applicable.
NOTE:
When performing these procedures, handle boards carefully and take the following precautionary
measures:
CAUTION:
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO
TAKING THE NECESSARY ESD PRECAUTIONS:
1.ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF
THE POWER CONNECTOR).
2.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
•
For BEP2 (RFT) and (RFI) systems
:
PC VIC Assembly -
see
.
ECG Box Assembly (Patient I/O Module) -
BEP2 - ECG Module Removal Procedure
.
Plug & Scan Battery -
see
Plug & Scan Battery Removal Procedure
.
PCB: Plug & Scan Card -
see
Plug & Scan Board Removal Procedure
.
PCB: SCSI Card -
see
.
MO Drive 5.25 -
.
Figure 8-41 BEP - Position of Brackets on BEP2 Models
Lower Right
Bracket
Lower Left
Bracket
Upper
Bracket