Section 8-5 - Back End Parts Replacement
8-5-12
MO Drive Replacement Procedure
8-5-12-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter.
8-5-12-2
Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in
.
8-5-12-3
MO Drive Removal Procedure
1) Disconnect the mains AC power cable for the mains power wall outlet.
2) Remove the Vivid™ 4 left and right side covers - refer to the
3.) Remove the Vivid™ 4 rear cover - refer to the
4) Remove the BEP left and right side covers, as described in the
5) Remove the BEP front cover as described in the
BEP2 - Front Cover Removal Procedure
6) Remove the cables (data and power) that connect to the rear of the MO Drive.
7.) Remove the two screws from each side of the MO that hold the MO Drive inside the BEP.
8.) Carefully slide the MO Drive out from the front of the BEP.
Figure 8-52 MO Drive Removal