GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-77
10.)Gently slide the power supply a short distance backwards (away from the BEP casing), while tilting
it downwards and pulling it to the side.
NOTE:
For clarification, the numbers appearing in square brackets in the following steps correspond to the
numbers shown in the illustration of the BEP2 Power Supply and cable connectors in
11.)Carefully remove the ECG Module from the front of the BEP, sliding it out and disconnecting the
power cable [
1
] from the rear of the ECG Module. (Refer to the
12.)Disconnect the power cables from the media drives (MOD [
2
] and CD-RW [
3
]).
13.)Remove the screw that fastens the Ground cable and PC-VIC in position, as shown in
14.)Disconnect the power cable from the Hard Disk (labelled [
4
] in
15.)Disconnect the data cable from the PC-VIC (refer to
16.)Slide the PC-VIC & hard disk out sufficiently to be able to cut any cable ties and disconnect the
power cable (labelled [
5
) from the PC-VIC; refer also to
Figure 8-61 BEP2 Power Supply with Cable Loom and Connectors
Figure 8-62 Disconnecting Cables
1
3
4
9
6
7
8
Not used
Cable Connections
1
ECG
2
MOD
3
CD-RW
4
Hard Disk
5
PC-VIC
6
Floppy Disk
7
BEP rear
8
Motherboard (beside ECG)
9
Motherboard (beside CPU)
2
5
Ground Cable Screw
Power Cable to PC-VIC
Data Cable to PC-VIC
Power Cable to Floppy Disk
Power Cable to Hard Disk
PC-VIC