Chapter 8 - Replacement Procedures
8-99
5) Loosen and remove the two M4 screws from the keyboard matrix (one on each of the
left
and
right
corners - see
below). Lift out the keyboard, completely removing it from the machine
chassis. For more details, refer to the
Note:
The up/down/swivel handle as well as the gas spring cable and swivel release cable are
now fully exposed, providing easy access to complete the handle removal procedure.
6) Using an M4 Allen key, loosen the two cap screws on the cable retaining bracket and slide the
bracket
forwards
sufficiently to release tension on the cables, as shown in
.
CAUTION:
When releasing tension, do not pull the cables excessively as this may cause damage
to the cable mechanism.
Figure 8-84 Removing the Keyboard
Figure 8-85 Gas Spring Cables and Retaining Bracket
Handle
Slide bracket in this direction
Loosen cap screws here
Gas spring cable
(up/down mechanism)
Swivel release cable
(swivel mechanism)
Cable retaining bracket