Chapter 10 - Periodic Maintenance
10-7
Section 10-5
System Periodic Maintenance
10-5-1
Preliminary Checks
The preliminary checks take approximately 15 minutes to perform. Refer to the
Vivid™ 4 User Manual
whenever necessary.
Table 10-4 System Preliminary Checks
St
ep
Item
Description
1
Ask & Listen
Ask the customer if they have any problems or questions about the equipment.
2
Paperwork
Fill in the appropriate details at the top of the
Vivid™ 4 Preventative Maintenance Inspection
(see page 10-31). Note all probes and system options.
3
Power-up
Turn the system power ON and verify that all fans and peripherals turn on. Watch the displays
during power up to verify that no warning or error messages are displayed.
4
Probes
Verify that the system properly recognizes all probes.
5
Displays
Verify proper display on the monitor and touch panel.
6
Presets
Back-up all customer presets on an CD-RW.