5.11 Air Flow Control
5.11.1 Air Flow Distribution
Through the filter grid on the back of the system (Main Air Inlet), air flow into the Voluson E-Series system.
•
Air holes in the RSP power supply allow the air to pass through; the 3 fans inside the RSP suck in the
air and spread it through the beamformer.
•
By means of the 2 Backend fans, air is blown through the GEB-box (along its internal components and
the PC- Motherboard).
The warm air exits the system through holes in the left side panel (Main Air Outlet).
Figure 5-17 Main Air Inlet/Outlet at Voluson E-Series
A Main Air Inlet (filter grid)
B Main Air Outlet
Components and Functions (Theory)
5-38
Voluson E-Series Service Manual
KTD106657 Revision 2