3.13.1 TCP/IP Configuration
Note
Following information must be provided by the customer or hospital engineer before you can start: Station
name, AE Title, IP address and Port Number for the Voluson E-Series. The IP addresses for the default
gateway and other routers at the site for ROUTING INFORMATION. Only if necessary (e.g. for Internet
access).
1.
Press the Utilities key on the control console.
2.
In the “Utilities” menu touch the
System Setup
button to invoke the setup desktop on the screen.
3.
On the left side of the screen select
Connectivity
and then click the
Device Setup
tab.
4.
Click the
TCP/IP Configuration
button, read the message and confirm with
Yes
.
Figure 3-65 TCP/IP Configuration
5.
The "Internet Protocol (TCP/IP) Properties" dialog page appears.
Note
This example shows fictional numbers!
Figure 3-66 Internet Protocol (TCP/IP)
1 You can get IP and DNS settings assigned automatically, if your network supports this capability.
2 If fixed IP address is required, ask your network administrator for the appropriate settings.
Type in:
-
IP address
-
Subnet mask
-
Default gateway
-
DNS server
Note
For further details refer to the Voluson E-Series Basic User Manual.
Setup Instructions
3-70
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...