4.2.2 Power Off / Shutdown
Note
After turning off a system, wait at least 10 seconds before turning it on again. The system may not be able to
boot if power is recycled too quickly.
1.
If not already in read mode, freeze the image.
2.
Press the ON/OFF Standby button on the control console. Following dialog appears.
Figure 4-3 Shutdown dialog
Figure 4-3 Shutdown dialog
3.
Select
Shutdown
. The system performs an automatic full shutdown sequence.
4.
Switch OFF the circuit breaker at the rear of the system.
Note
A full shutdown is also performed when pressing the ON/OFF standby button on the control console twice.
Note
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the ON/OFF
standby button. So the auxiliary equipment need not to be switched ON/OFF separately.
Warning
Disconnection of the main power cable is necessary!
5.
After complete power down, unscrew the 2 screws and remove the pull-out protection to disconnect the
main power cable from the system or unplug it from the AC wall outlet socket.
Figure 4-4 Circuit Breaker at rear of system
1 circuit breaker
2 fuses (2x T10A H/250V)
3 connector for main power cable
6.
Press on the brakes to block the front caster wheels.
7.
Disconnect probes. (Turn the probe locking handle counterclockwise and then pull the connector
straight out of the probe port.)
Caution
Do not disconnect a probe while running (Live Scan “Write” mode)! A software error may occur. In this case
switch the system OFF (perform a reset).
Functional Checks
4-4
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...