Step
Task
Expected Results
29
Line Filter
The signals of the neighboring pulses are less weighted for the display of the actual
pulse which considerably improves the detail lateral resolution and signal-to-noise ratio.
This function is only available if
XBeam CRI is switched off.
30
Line Density
Control to improve the resolution by reducing the frame rate. Respectively reducing the
resolution by increasing the image frame rate.
31
Enhance
Edge Enhance brings out subtle tissue differences and boundaries by enhancing the
gray scale differences corresponding to the edges of structures.
32
Reject
Selects a level below which echoes will not be amplified (an echo must have a certain
minimum amplitude before it will be processed).
Note
For further details refer to the Voluson E-Series Basic User Manual.
4.3.2 Additional (optional) Operating Modes
4.3.2.1 B-Flow
B-Flow On/Off switch
Press the BF key to activate/deactivate the B-Flow mode.
Figure 4-9 B-Flow Main and B-Flow Sub Menu
Note
For further details refer to the Voluson E-Series Basic User Manual.
Functional Checks
Voluson E-Series Service Manual
KTD106657 Revision 2
4-11
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...