4.3.6 Volume Mode
Note
4D Real Time 4D, RT_4D_Biopsy, VOCAL II, etc. are Options. If options are not part of the system
configuration, checks can be omitted.
Note
Different menus are displayed depending on which Touch Panel menu and which Volume Mode is selected.
Note
Some function keys only appear on the Touch Panel if they are available for the selected probe.
4.3.6.1 Pre-Volume Mode Functions
1. Press the 3D or 4D key to select the desired Volume Mode.
The Volume Mode function is switched on, the “3D Pre” respectively “4D Pre” menu appears on the screen
(write mode) and the volume box appears on the Image area.
Figure 4-16 Pre-Volume Mode menu (e.g., 4D Real Time)
2. Select the Acquisition Mode (2).
3. Select an User Setting (3).
4. Select the desired Visualization Mode (4).
5. Start the Volume Acquisition with the Freeze key resp. the right trackball key.
Table 4-5 Pre-Volume Mode Functions
Step
Task
Expected Results
1
ACQUISITION:
3D Static
3D Volume Mode - Static volume acquisition (also in combination with PD, HD-Flow or
CFM)
4D Real Time
Real Time 4D - continuous volume acquisition and parallel calculation of 3D rendered
images
VCI A
Volume Contrast Imaging - improves the contrast resolution and the signal / noise ratio
and therefore facilitates finding of diffuse lesions in organs
VCI Omni View
Volume Contrast Imaging Omni View (any plane) - improves the contrast resolution and
the signal / noise ratio and therefore facilitates finding of diffuse lesions in organs.
Volumes can be edited in all other Visualization Modes.
STIC
The fetal heart or an artery can be visualized in 4D (also in combination with PD, HD-
Flow or CFM)
Functional Checks
Voluson E-Series Service Manual
KTD106657 Revision 2
4-17
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...