5.2.1 RTF - Probe Control Board
Switches the Probe Connectors (3 DLP-Connectors, 1 CW-Connector) and recognizes Probes
•
1 CW-Probe Connector
•
3 Probe-Connectors 408pin
•
1 Dummy-Probe Connector 408pin
•
Probe Select Relays
•
Probe Recognition
5.2.2 RSE - Pencil Probe Board (optional)
Adapter board for connection of CW pencil probes.
Note
The RSE board is required for CW-Option.
5.2.3 RFM - (RF-Interface & Beamformer) FE Mainboard
The FrontEnd Mainboard supports Tx/Rx for 128 channels only. To extend to 192 or 256 channels, the
- (Beamformer Receiver/Transmitter) Extension
5.2.3.1 RFM Board - Interface FPGA
1.
DMA logic
2.
Beamformer Interface
3.
RTF Control Interface
4.
RTF FPGA Control Interface
5.2.3.2 RFM Board - Processing FPGA
1.
Ultrasound Data Pre-Processing
2.
System Control
3.
Motor Control
5.2.4 RSX - (Beamformer Receiver/Transmitter) Extension Board
Subset (for RFM) that is required to extend to 192 or 256 channels.
Note
All components of RSX board are also present on
RFM - (RF-Interface & Beamformer) FE Mainboard
.
Components and Functions (Theory)
Voluson E-Series Service Manual
KTD106657 Revision 2
5-21
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...