7.5.2 How the Customer enables/disables Disruptive Mode and VCO
1.
If not already in read mode, Freeze the image.
2.
Move the cursor to the InSite ExC link (
GE
icon at the right bottom of the display screen) and press the
right trackball key (= right-click).
3.
Select
Connect Clinical Lifeline
). This activates “Disruptive Mode” and “VCO”
for the application OLC to quickly assist the customer.
Figure 7-11 Connect Clinical Lifeline
4.
If you select
Cancel
, “Disruptive Mode” and “VCO” is turned OFF.
Note
Visual information about GE remote service status is shown in the status area on the right bottom of the
screen.
Gray = Idle State
Remote access is inactive.
Yellow = Access pending
Remote Access connection is active, but Disruptive Mode and Virtual Console
Observation (VCO) is not yet enabled.
Red = Disrupted State
Remote Access is active. All processes [UL_VNC and UL_CSD] are active. In
this state the Voluson E-Series system should NOT be used clinically.
Gray + envelope
New Software Update Package is available for download and installation.
Operation see
Section 8.2.3 "Software Update Package - Download/
Installation Procedure" on page 8-9
If a GE Service Technician requests Remote Access Permission
If a GE Service technician requests remote access to your Voluson E-Series system, following message
appears on the screen.
Figure 7-12 Insite User Session Message
To enable Disruptive Mode click
OK
.
Note
If the customer does not wish to have diagnostics running at the time of the request, select Cancel. A
message is sent back to the OLC or FE that permission to change Disruptive Status is denied.
Diagnostics/Troubleshooting
7-12
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...