7.7 How to use the Auto Tester program
1.
Press the Utilities key on the control console.
2.
In the “Utilities” menu touch the
System Setup
button to invoke the setup desktop on the screen.
3.
On the left side of the screen select
Administration
and then click the
Service
tab.
4.
Enter the password <SHE> and click the
Accept
button to display the Service Tools page (see
).
5.
Activate the “Auto Tester” program by clicking
Start
. The following message box appears.
Figure 7-22 Autotester activation
6.
Click
OK
.
7.
Press the Alt Gr key on the alphanumeric keyboard.
8.
Activate the “Auto Tester” program by clicking the
“Record”
icon on the displayed screen.
Figure 7-23 Autotester record
9.
Start scanning. You can scan normally and everything will be recorded to the program (up to several
hours).
Note
It is important that you are recording the processes where the errors normally occur.
Stop the program by clicking on [Stop] shown on the screen, or by pressing the Alt Gr key
on the alphanumeric keyboard.
Diagnostics/Troubleshooting
Voluson E-Series Service Manual
KTD106657 Revision 2
7-17
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...