Item
Service at indicated Time
D W M A
Notes
Measurement Accuracy
Checks
•
Also after corrective maintenance or
as required by your facilities QA
program.
Probe/Phantom Checks
Check gain and TGC changes, vary the focus and
check reaction on screen.
NOTE! The use of a Phantom is not required during
Preventive Maintenance. Customer may use it as part
of their Quality Assurance Program tests.
•
Also after corrective maintenance or
as required by your facilities QA
program.
Functional Checks of all
probes
•
Also after corrective maintenance or
as required by your facilities QA
program.
10.3 Tools required
10.3.1 Special Tools, Supplies and Equipment used for Maintenance
Note
Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool compliant
with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551 or IEC 62353 or other relevant national regulation.
Table 10-2 overview of supplies and equipment used for maintenance
Tools
Comments
Digital Volt Meter (DVM)
minimum 5% accuracy, 3.5 digit and 200 Ohm range required
Anti Static Kit
including anti-static mat, wrist strap, cables and conductive ground cord
Anti Static Vacuum Cleaner
•
QIQ Phantom
NOTE! The use of a Phantom is not required during Preventive Maintenance. Customer
may use it as part of their Quality Assurance Program tests.
CD-RW Media
(minimum quad speed)
DVD+RW Disc Media blank
blank 4,7GB DVD+RW disc
B/W Printer Cleaning Sheet
see printer user manual for requirements
Color Printer Cleaning Sheet
see printer user manual for requirements
Disposable Cloves
•
Care and Maintenance
10-4
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...