10.4.3 Physical Inspection
Table 10-7 physical checks
Step Item
Description
1
Labeling
Verify that all ultrasound system labeling is present and in readable condition.
2
Covers
Where applicable, verify all covers are secured in place and are properly aligned with other covers.
Replace any covers that are damaged.
3
Control Console
and Keyboard
Inspect the control console and keyboard. Record any damaged or missing items. (Replace faulty
components, as required).
4
Control Console
Movement
• Where applicable, verify ease of control console movement in all acceptable directions.
• Where applicable, ensure that it latches in position as required.
5
Control Console
Lighting
Check proper operation of all control console and TGC lights.
6
LCD Monitor
Inspect the LCD Display for scratches and bad pixels. Verify proper operation of Contrast and
Brightness controls. Where applicable, confirm that the LCD arm allows:
• swivelling the screen to the left and to the right
• folding the screen to the locked position
• release and adjustment backwards and forwards
• can be adjusted in the up/down positions
Note: LCD Arm movement may vary and is not applicable to all ultrasound systems.
7
Monitor Light
Check for proper operation of any monitor lights, if available.
8
Cables and
Connectors
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
special attention to footswitch assembly and probe strain or bend reliefs.
9
Shielding & Covers Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems during scanning.
10
Probe Holders
Where applicable, inspect the Probe Holders for cracks or damage.
11
Power and System
Status Indicators
Check for proper operation of all Power and System Status Indicators.
12
External I/O
Check all connectors for damage and verify that the labeling is good.
13
Wheels and Brakes • Check all wheels and casters for wear and verify operation of foot brake, to stop the system from
moving, and release mechanism.
• Check all wheel locks and wheel swivel locks for proper operation.
14
Battery
Where applicable, check that the battery is not damaged, does not leak, does not emit an odor, and is
not deformed or discolored. Observe all warnings and cautions for battery handling, recharging,
storing, and/or disposal.
Care and Maintenance
Voluson E-Series Service Manual
KTD106657 Revision 2
10-7
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...