10.6.3 Outlet Test - Wiring Arrangement - USA & Canada
Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester
and noting the combination of lights that are illuminated. Any problems found should be reported to the
hospital immediately and the receptacle should not be used.
Figure 10-1 Typical Alternate Outlet Tester
1
Correct Wiring
3
Reversed Polarity
5
Hot and Ground Reversed
2
Open Ground Wire
4
Open Neutral Wire
6
Open Hot Wire
Note
No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the Grounding
(protective earth) conductor are reversed. If later tests indicate high leakage currents, this should be
suspected as a possible cause and the outlet wiring should be visually inspected.
Care and Maintenance
Voluson E-Series Service Manual
KTD106657 Revision 2
10-13