3.4.10 Connecting a Secondary "Patient" Monitor
Caution
A Secondary "Patient" Monitor MUST NEVER be connected to the Voluson E-Series systems mains supply
directly! Always connect it to an appropriate Isolation Transformer (see
).
Caution
The secondary monitor is the only item to be connected to the isolation transformer.
Note
A Secondary Monitor is NOT intended for diagnostic use. It is an additional device used to allow the patient
to watch the proceedings.
Note
Take your time to think about the best position of the monitor in your facilities. Patients should be able to
view the monitor easily and without having to bend or turn around.
1.
Power Off/Shutdown the system as described in
2.
Connect the Secondary Monitor according to connection scheme, see:
.
3.
After physical connection to the Voluson E-Series system check, and if necessary change the Ext.
Monitor setting, see:
Note
Connection is always the same (no differences between PC-Motherboard version of the Voluson E-Series
system).
Setup Instructions
3-32
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...