3.5 Completing the Setup
Connecting the System to a Power Source
Caution
Prior to connect the Voluson E-Series system to a power source, verify compliance with all electrical and
safety requirements. Check the power cord to verify that it is intact and of hospital-grade. Products equipped
with a power source (wall outlet) plug should be connected to the fixed power socket that has a protective
grounding conductor. Never use an adapter or converter to connect with a power source plug (for example, a
three-prong to two-prong converter).
Warning
The system's power must be supplied from a separate, properly rated outlet to avoid risk of fire. Refer to
Section 2.1.2.1 "Voluson Power Requirements Voluson E-Series" on page 2-2
power cord should not, under any circumstances, be altered to a configuration rated less than that specified
for the current.
Note
Use only the power cords, cables and plugs provided by or designated by GE Healthcare Austria GmbH &
Co OG to connect the system to the power source.
Caution
Whenever disconnecting the Voluson E-Series system from the electrical outlet, always observe the safety
precautions. First unplug the main power cable from the wall outlet socket, then from the system itself.
Remove by pulling on the cable connector - Do not pull on the cable.
Caution: The Voluson E-Series requires all covers!
Do not operate this system unless all board covers and frame panels are securely in place, to ensure optimal
system performance and cooling. (When covers are removed, EMI may be present).
3.5.1 Power On / Boot Up
System Power On / BackEnd Processor Boot Up
1.
Connect the main power cable to the back of the system.
2.
If not already done, screw on the pull-out protection of the mains power cable with the 2 screws.
3.
Connect the main power cable to a hospital grade power outlet with the proper rated voltage. Never
use an adapter that would defeat the safety ground.
4.
Switch ON the circuit breaker at the rear of the system.
Figure 3-29 Circuit Breaker at rear of system
1 circuit breaker
2 fuses (2x T10A H/250V)
3 connector for main power cable
Note
When AC power is applied to the system, the ON/OFF standby button on the control console illuminates
amber, indicating that the system (including the Back-end Processor) is in standby mode.
Setup Instructions
Voluson E-Series Service Manual
KTD106657 Revision 2
3-39
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...