September 2007
26
HygroTrace Moisture Transmitter
Specifications
General
Measurement Range:
•
0 to 100 ppb
v
with trend indication beyond the calibrated range
Compatible Gases:
•
Nitrogen and Argon
Process Gas Temperature Range:
•
14°F to 95°F (–10°C to 35°C)
Storage Temperature:
•
–40°F to 158°F (–40°C to 70°C)
Operating Temperature:
•
14°F to 95°F (–10°C to 35°C)
Warm-Up Time:
•
Meets specified accuracy within 24 hours, after sensor
exposure <72 hours @ 25°C and 60% RH
Calibrated Uncertainty @ 77°F (25°C):
•
±20% of reading or ±5 ppb
v
, whichever is greater
Response Time:
•
Less than 20 min. for 95% of 25 ppb
v