Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5
pim 2337741–100
35
1-3-3
ELECTROMAGNETIC IMMUNITY
Table 17 – Electromagnetic Immunity
The Innova 2000 system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Innova 2000 system should assure that it is used in an
electromagnetic environment as described below:
Immunity Test
IEC 60601–1–2
Test Level
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000–4–2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors are wood, concrete, or ceramic tile,
or floors are covered with synthetic material
and the relative humidity is at least 30
percent.
Electrical fast
transient/burst
IEC 61000–4–4
2 kV for power
supply lines
1 kV for
input/output
lines
2 kV for power
supply lines
1 kV for
input/output
lines
Mains power quality is that of a typical
commercial and/or hospital environment
Surge
IEC 61000–4–5
1 kV line(s) to
lines(s)
2 kV line(s) to
earth
1 kV line(s) to
lines(s)
2 kV line(s) to
earth
Mains power quality is that of a typical
commercial and/or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000–4–11
0 % U
n
for 5 sec
0 % U
n
for 5
sec
Mains power quality is that of a typical
commercial and/or hospital environment. If
the user of the Innova 2000 system
requires
continued operation during power mains
interruptions, it is recommended that the
Innova 2000 system
be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000–4–8
3 A/m
3 A/m
1A/m
Power frequency magnetic fields are at
levels characteristic of a typical location in a
typical commercial and/or hospital
environment.
At that disturbance amplitude level the
monitor image might present some slight
flicker. For image quality improvement, the
monitor might be removed from the low
frequency magnetic field source vicinity.
At that disturbance amplitude level the
system is fully operational.
Note: These are guidelines. Actual conditions may vary.