2.16 Omega V Long Table
Illustration 9-34: Omega V long table - Slab on grade 1/2
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
242
2 Calculations
Page 1: ...GE Healthcare Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual OPERATING DOCUMENTATION 5177951 4 1EN Revision 1...
Page 2: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 2...
Page 3: ...Important Information LANGUAGE Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Important Information 3...
Page 4: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 4 Important Information...
Page 5: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Important Information 5...
Page 6: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 6 Important Information...
Page 7: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Important Information 7...
Page 8: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 8 Important Information...
Page 9: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Important Information 9...
Page 10: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 10 Important Information...
Page 11: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Important Information 11...
Page 12: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 12 Important Information...
Page 13: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Important Information 13...
Page 14: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 14 Important Information...
Page 15: ...r Change Pages 5177951 4 1E N rev 1 25 March 2010 Initial release of 5177951 4 1EN 252 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177...
Page 16: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 16 Revision History...
Page 17: ...workstation option 32 1 9 CENTRICITY CA1000 option 32 1 10 Injectors 32 1 11 Fluoro UPS Option 32 1 12 Touch Screen 35 1 13 IntraVascular UltraSound IVUS Option 35 1 13 1 Application 35 1 13 2 Purpos...
Page 18: ...patibility Cross Reference Monitor Support Suspension Sub System 51 3 7 System Compatibilities Cross Reference 18 19 LCD Monitors Sub System 51 Chapter 3 PHYSICAL REQUIREMENTS OF ROOM 53 1 Presentatio...
Page 19: ...2 6 Windows and curtains 65 3 Structural Requirements 66 3 1 Room Size 66 3 2 Elevator and Door Size Requirements 66 3 2 1 Door Height 66 3 2 2 Door Width 66 3 2 3 Elevator Depth 66 3 3 Floor 66 3 3 1...
Page 20: ...hapter 4 ROOM LAYOUTS 85 1 Room Layout Considerations 85 1 1 Service Access 85 1 2 Clinical Access 85 1 3 Peripheral Equipment 85 1 4 Emergency Stop 85 1 5 Patient Environment Equipment 86 1 6 Recomme...
Page 21: ...h insert S18101SF 140 2 3 3 6 2285632 Floor plate Base plate assembly S18101SG 141 2 3 4 2285056 A C N Templates 141 2 4 Injector Mounting Requirements 144 Chapter 6 ELECTRICAL CONNECTIONS 145 1 Power...
Page 22: ...e Windows 174 15 MAC LAB System EX 175 16 MIS Master Interconnect System 176 17 Cable Channeling 177 17 1 General 177 17 2 Conduit 177 17 3 Electrical Ducts 177 18 Physical Runs 180 Chapter 7 ADDITION...
Page 23: ...Chapter 8 IP ADDRESSING PROCESS 205 1 IP Addressing Process 205 Chapter 9 SEISMIC CALCULATIONS 207 1 Overview 207 2 Calculations 208 2 1 C1 Frontal Cabinet 209 2 2 C1 Lateral Cabinet 210 2 3 C2 LC LP...
Page 24: ...238 2 16 Omega V Long Table 242 2 17 Overhead Flat Panel Suspension Ceiling mounted 246 2 18 8 Monitor Flat panel Suspension Ceiling Mounted 248 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging...
Page 25: ...e Installation Specialist and the Site Planner in properly preparing a site for the installation of this system It provides pre installation data such as site preparation prior to the delivery of the...
Page 26: ...he customer that the cost of any alteration or modification not specified in the sales contract are liable to the customer The following GE supplied equipment must be installed by the Hospital s Contr...
Page 27: ...ss The checklists summarize the required preparations and allow to verify the proper completion of the pre installation procedures You will find hereafter a chart of the information flow in the pre in...
Page 28: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 28 3 Pre Installation Process...
Page 29: ...or and a Revolution Digital Detector Two C1 cabinets including X Ray Generator and Digital Imaging System One C2 LC LP cabinet including positioners control One or two multi axis control boxes 1 2 Inn...
Page 30: ...y tube housing Two External recirculating chillers mandatory for each tube It is mandatory to place the two Chillers in the Technical Room 1 4 X Ray head The Innova System uses two Ulysses X Ray tube...
Page 31: ...W Monitor LCD 19 EIZO GmbH SMD 19100G B W Monitor 1 6 Monitor suspensions The common type of this suspension is an XT inboard monitor bridge A monitor frame support receiving 6 monitors 4 B W Monitor...
Page 32: ...jectors The recommended injectors are ILLUMINA pedestal ACIST E2000 Voyager pedestal ACIST CVI Pedestal and Table mount Mark V ProVis pedestal or rack mounted MEDRAD AVANTA Pedestal and Table mount Ot...
Page 33: ...able parts inside All installation maintenance and service work should be performed by qualified service personnel The UPS contains its own energy source battery The field wiring terminals may be elec...
Page 34: ...object into the UPS The unit must be placed in a sufficiently ventilated area the ambient temperature should not exceed 104 F 40 C Optimal battery life time is obtained if the ambient temperature doe...
Page 35: ...for IntraVascular UltraSound imaging designed as add on equipment to Vascular Imaging systems The IVUS s5i GE equipment is also capable of recieving patient data from the Innova system The Volcano IVU...
Page 36: ...onal Flow Reserve mode which serves as a blood pressure measurement device in the vascular veins In this mode the Volcano s5x device can measure the difference between the patient s aortic and distal...
Page 37: ...ioner L arm Item 1 Pivot Item 2 C arc Item 3 Motorized elevator Item 4 for the Revolution Digital Detector X ray tube Item 5 Collimator Item 6 21 cm Revolution Digital Detector for Innova 2121 IQ Item...
Page 38: ...12 Collimator Item 13 21 cm Revolution Digital Detector for Innova 2121 IQ Item 14 31 cm Revolution Digital Detector for Innova 3131 IQ Item 14 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging S...
Page 39: ...1 Lateral Cabinet Illustration 2 3 C1 frontal and lateral cabinets Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1...
Page 40: ...binet Illustration 2 4 C2 LC LP Cabinet Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 40 2 Innova LC LP Product...
Page 41: ...ion 2 5 Omega tables 2 5 Tilt Patient Table Illustration 2 6 Tilt table Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revis...
Page 42: ...e Illustration 2 7 User interface devices Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 42 2 Innova LC LP Produc...
Page 43: ...Fluoro UPS option Illustration 2 8 UPS UL Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 2 Innova System...
Page 44: ...Illustration 2 9 UPS CE Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 44 2 Innova LC LP Product Identification...
Page 45: ...tion 2 10 3 kVA UPS model 9125 Illustration 2 11 3 kVA UPS model 9130 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revisio...
Page 46: ...Lateral 2 10 Digital Detector Conditioner Illustration 2 13 Thermo con chillers Frontal and Lateral Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare D...
Page 47: ...ser Area Illustration 2 14 Innova User Area devices Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 2 Inno...
Page 48: ...Illustration 2 15 IVUS Rev 1 components Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 48 2 Innova LC LP Product...
Page 49: ...Illustration 2 16 IVUS Rev 2 components Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 2 Innova System 49...
Page 50: ...s Omega V Long Table This document Includes tabletop Omega IV Table This document Includes tabletop 3 3 System Compatibilities Cross Reference Jedi X Ray Generator Sub System Product Name Pre Installa...
Page 51: ...duct Name Pre Installation Document Number Catalog Number Precabled LCD 6 or 8 Monitor Suspension Cable Harness 24 M 78 ft 9 in or Cable Harness 36 m 118 ft 2 in 2393190 100 3 7 System Compatibilities...
Page 52: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 52 3 System Compatibility...
Page 53: ...E IS A RISK OF TRANSMISSION OF THESE AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL PERSONNEL NOSOCOMIAL DISEASES TO REDUCE THIS RISK THE ELECTRONIC CABINETS MUST BE INSTALLED IN A ROO...
Page 54: ...C 59 F 32 C 90 F Design for Patient Operator Comfort 10 C 14 F 50 C 122 F Technical room See Note 1 30 75 10 70 10 C 50 F 35 C 95 F 13 C 55 F 25 C 77 F 14 C 8 F 50 C 122 F Technical room with Fluoro...
Page 55: ...or chiller HEAT DRY 1 is shipped within GE Healthcare packaging NOTE 3a In use temperature limits specify the range where the system shall work Operating outside these limits could occur severe perfor...
Page 56: ...g toTable 3 1 and Table 3 2 2 2 Equipment Heat Output Dissipation 2 2 1 Standard System Refer to Table 3 3 To obtain heat output information for components not specified in Table 3 3 refer to the appr...
Page 57: ...1 709 0 21 709 0 21 709 0 21 709 Main disconnect panel PDB 0 60 2216 0 65 2216 0 65 2216 0 65 2216 3kVA Cabinet UPS mod el 9125 0 16 546 0 16 546 0 16 546 0 16 546 3kVA Cabinet UPS mod el 9130 4 0 37...
Page 58: ...inlet grids located at the bottom and exhausted through the outlet on the roof A suitable ventilation or cooling system must be installed to extract the heat from the UPS room NOTICE Do not put anyth...
Page 59: ...0 80 Temperature at 20 Relative Humidity 5 35 C 41 95 F Altitude Relative to see level max 3048 m max 10000 ft 2 2 3 2 Heat Output Device Location Heat output PC Control room 335 W 1206 BTU h Monitor...
Page 60: ...em is suitable for use only in establishments not directly connected to a public low voltage power supply network Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable The Innova system...
Page 61: ...urchaser or user of the Innova system should assure that it is used in an electromagnetic environment as described below Table 3 9 Immunity Test IEC 60601 1 2 Test Level Compliance Level Electromagnet...
Page 62: ...may be necessary such as re orienting or relocating the Innova system Over the frequency range 150 kHz to 2 5GHz the recommended field strengths is here less than 0 3V m The Recommended Separation Dis...
Page 63: ...d store diagnostic quality images safely For example a 1 W mobile phone 800 MHz to 2 5 GHz carrier frequency shall be put 23 meters see Table 3 10 apart from the Innova system in order to avoid images...
Page 64: ...omponents accessories subsystems and options as the Advantage Workstation and the UPS That same AC power distribution panel should be supplied by a separated AC power line coming from a separated tran...
Page 65: ...ined inTable 3 11 are the results of the measurement performed on the first Innova 3131 IQ system installed on a clinical site Table 3 11 Gantry Noise measurements Measurements Gantry FRT LAT both Gan...
Page 66: ...3 2 3 Elevator Depth The minimum elevator depth needed to accommodate the Innova LP4 shipping dolly is 2954 mm 116 3 in 3 3 Floor 3 3 1 General Vascular GE Healthcare Policy GE Healthcare s Customer...
Page 67: ...rength of 17 24 MPa at 28 days which is the minimum required compression strength Anchors installed to the required hole depth of 165 1 mm 6 5 in minimum and Center of anchor hole to concrete edge dis...
Page 68: ...s Pre Installation Kit Installation Procedures Illustration 3 4 Through Bolt Supplied Pan Type Floor Construction Illustration 3 5 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Inst...
Page 69: ...not placed on a computer floor but directly put on concrete floor the location of the cable access needs to be carefully planned Otherwise if the cable run is located under the concrete floor the cabl...
Page 70: ...ration 3 6 Hole location in concrete floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 70 3 Structural Require...
Page 71: ...it plus bushing is limited to 12 mm 1 2 in NOTE Refer to table Chemical anchors Pull out efforts and recommendations in Chapter 5 Mounting Requirements for pull out effort on each fixation bolts 3 3 5...
Page 72: ...efer to Innova LC Positioner and Omega Tilt Table Floor Preparation Kits GE Healthcare supplied in Chapter 5 Mounting Requirements Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Inst...
Page 73: ...he minimum curvature constraint of 150 mm 6 the cable will have to come from the side between Innova LC LP Positioner and patient table Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre...
Page 74: ...1 in The recommended distance is equal to 4 275 m 14 ft This includes the 150 mm 6 in clearance between the end of the rails and the wall SeeIllustration 3 11 andIllustration 3 12 Potential Wall Inter...
Page 75: ...the maximum configuration Illustration 3 11 Some room configurations may require shorter configurations Illustration 3 12 shows the Absolute Minimum configuration allowed Innova 2121 IQ 3131 IQ Bipla...
Page 76: ...on of Stationary Rails on Ceiling Absolute min configuration Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 76 3...
Page 77: ...13 LP4 Rail Mounting Specifications Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 3 Physical Requiremen...
Page 78: ...pport structure must be leveled before installation can begin Do not assume that any support structure is level within specified tolerances particularly after removing suspensions from an existing roo...
Page 79: ...kVA xxx 480 V self transformer for chillers 120 180 60 Japan 380 to 460 10 243 per C1 cabinet 23 per chiller 20 for C2 cabinet 19 to 17 3Ph PG A 150kVA xxx 480V self transformer NOTE About X Ray tube...
Page 80: ...rating is equal to 211 kVA NOTICE Line impedance should be compliant with IEC 601 2 7 Refer to the table Max Line Impedance for feeder line between Generator cabinet and Hospital in Chapter 6 Power a...
Page 81: ...ces or by reflection on bright objects The Illumination level must be compliant with established lighting technical rules and be as constant as possible Technical room operating room and control room...
Page 82: ...nding to assembly of medium size electrical components e g control panel for the electrical industry Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare...
Page 83: ...t to finish the examination In this case a wall outlet single phase ground is required to feed the mobile It is also requires a free space around the patient table to proceed with the mobile instead o...
Page 84: ...nection to a modem will preferably be located close to the cabinet The phone outlet must be located less than 1 meter 3 feet from the C1 cabinet A modem compliant to each country is supplied with the...
Page 85: ...There must be an unrestricted view of the video monitors and physiological monitoring equipment from the vascular table Operators in the control area must have easy access to the control console Howev...
Page 86: ...equipment patient vicinity is defined in the standardization as a space within the room 1 83 m 6 ft beyond the perimeter of the examination table and extending vertically 2 29 m 7 ft 6 in above the f...
Page 87: ...eiling height top surface of LP4 stationary rails such as Unistrut mounting bolts support brackets sprinklers air vents etc 1 7 Preference Cabinet locations NOTE This is applicable for all types of ca...
Page 88: ...straints Technical and Exam room on same floor Location Constraint Name Parameter Specification Comment Technical room Cabinets location Maximum relative distance 9 m 29 ft 6 in Make sure to provide e...
Page 89: ...cification Comment Technical room Cabinets location Maximum relative distance 9 m 29 ft 6 in Make sure to provide enough space for storing extra cable length e g Plinth Technical room Tube chillers lo...
Page 90: ...tion and equipment and the axis of the equipment L axis versus the remote controls dead angles concern WARNING CARRIAGE COVERS CAN ENCLOSE DUST PARTICLES CARE SHOULD BE TAKEN TO AVOID PROPAGATION IT I...
Page 91: ...38 ft 0 in x 26 ft 2 in 2845 mm 5 mm 9 ft 4 in 0 2 in is mandatory ceiling height Minimum 6900 mm x 4400 mm 22 ft 8 in x 14 ft 5 in Table 4 4 Room Layout components see Illustration 4 3 1 3 kVA UPS ca...
Page 92: ...twork drop outlet must be located less than 1 meter 3 feet from the C1 Frontal Cabinet 20 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5...
Page 93: ...ustration 4 3 Room Layout for Innova System Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 4 Room Layouts...
Page 94: ...ption Room Layout Illustration 4 4 IVUS Rev 1 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 94 3 IVUS Option Roo...
Page 95: ...Illustration 4 5 IVUS Rev 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 4 Room Layouts 95...
Page 96: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 96 3 IVUS Option Room Layout...
Page 97: ...Omega Table side clearence CPR access Illustration 5 9 Gas Box Outlets Omega Table 1 Illustration 5 10 Innova LC LP Positioner And Omega Patient Table Relative Positions Side View 1 Illustration 5 11...
Page 98: ...Dimensions Side View Illustration 5 2 Innova LC Positioner Dimensions Top View Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1...
Page 99: ...Innova LC Positioner Dimensions Front View Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Cha...
Page 100: ...ion 5 4 Innova LP4 Positioner Dimensions Side View Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 100 1 Dimension...
Page 101: ...Innova LP4 Positioner Dimensions Top View Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Char...
Page 102: ...on 5 6 Innova LP4 Positioner Dimensions Front View Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 102 1 Dimension...
Page 103: ...tion 5 7 Omega IV Table Dimensions Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Characterist...
Page 104: ...end rail If there are any devices installed on the table foot end the width of these devices must be added to the existing 500 mm 19 7 in to maintain absolute minimum distance of 500 mm 19 7 in Innova...
Page 105: ...ation 5 8 Omega V Table Dimensions Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Characterist...
Page 106: ...end rail If there are any devices installed on the table foot end the width of these devices must be added to the existing 500 mm 19 7 in to maintain absolute minimum distance of 500 mm 19 7 in Innova...
Page 107: ...9 Omega Table side clearence CPR access Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Charac...
Page 108: ...lustration 5 10 Gas box outlets Omega table Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 108 1 Dimension Drawin...
Page 109: ...e 5 2 2 in the case of LP4 Off iso feature installed the LP4 will work in the range 200 mm 7 9 in and 200 mm 7 9 in from isocenter 3 LP4 parking not allowed at patient foot end Innova 2121 IQ 3131 IQ...
Page 110: ...r distances ANGIO CARDIO CARDIO NEURO Omega IV NA 1395 mm 54 9 in Omega V non motorized 1278 mm 50 3 in 1395 mm 54 9 in Omega V motorized 1278 mm 50 3 in 1395 mm 54 9 in Innova 2121 IQ 3131 IQ Biplane...
Page 111: ...is includes sag between drape points A B C and D The worst case LP4 cable drape extension including sag is 1 85 m 72 8 in between A and B 2 05 m 80 7 in between B and C 2 30 m 90 5 in between C and D...
Page 112: ...lustration 5 14 C2 LC LP Cabinet Dimensions Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 112 1 Dimension Drawin...
Page 113: ...on 5 15 C1 Frontal Cabinet Dimensions Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Character...
Page 114: ...ustration 5 16 C1 Lateral Cabinet Dimensions Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 114 1 Dimension Drawi...
Page 115: ...00 chiller depend on model type Required floor space depends on ambient room temperatures When in doubt allow for maximum floor space Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre I...
Page 116: ...rmo Con Detector Conditioner and Mounting brackets Dimensions Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 116...
Page 117: ...on 5 19 3 kVA Cabinets UPS model 9125 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Character...
Page 118: ...ustration 5 20 3 kVA Cabinets UPS model 9130 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 118 1 Dimension Drawi...
Page 119: ...area to insure electrical safety shall be at least 0 9 m In cases where 2 cabinets are installed face to face both sides of the access way the clear width shall be at least 1 2 m Recommended minimum...
Page 120: ...intenance operation In front of the cabinet the clear width of the service area to insure electrical safety shall be at least 0 9 m In cases where 2 cabinets are installed face to face both sides of t...
Page 121: ...ration 5 23 DL Keypad Dimensions Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Characteristic...
Page 122: ...lustration 5 24 DL Image Monitor Dimensions Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 122 1 Dimension Drawin...
Page 123: ...Illustration 5 25 VCIM Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Characteristics 123...
Page 124: ...e 754 6 1664 See note 1 See Illustration Omega Table Di mensions in Dimension Drawings 571 5x429 22 5x16 9 2410 kg m2 492 3 lb ft2 Same as Innova LC Positioner C2 LC LP Cabinet 283 624 See Illustratio...
Page 125: ...Rev 2 Option IVUS PIM Ultra sound probe 2 4 4 38 96 177 IVUS Control Sta tion 4 8 8 120 360 240 IVUS Joystick 1 2 2 90 76 102 IVUS Isolation transformer 7 15 4 97 267 171 IVUS CPU 15 33 1 396 162 510...
Page 126: ...floor mounting templates to ensure these components are properly placed in relation to one another Table 5 4 Title Illustration Innova LC Positioner Floor Mounting Methods Illustration 5 26 and Illust...
Page 127: ...nnova LC Positioner Floor Mounting Methods 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical...
Page 128: ...27 Innova LC Positioner Floor Mounting Methods 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 128 2 Mounting...
Page 129: ...Cable Conduit For On Grade Floor Anchor Kit Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Ch...
Page 130: ...29 Inner Base Plate For Above Grade Floor Anchor Kit Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 130 2 Mountin...
Page 131: ...E For more details on Omega Tilt Baseplate refer to Illustration 5 31 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revisio...
Page 132: ...per bolt with 4 bolts 272 daN per bolt if 24 used and 2008 daN per bolt if 12 used Number of holes in the plate 12 max 8 min mandato ry 10 max 8 min mandato ry 4 mandatory 24 max 12 min mandato ry Re...
Page 133: ...n 5 31 Gantry and table mounting holes Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Characte...
Page 134: ...NA 1395 mm 54 9 in Omega V Long 1278 mm 50 3 in 1395 mm 54 9 in Omega V non motorized Long 1278 mm 50 3 in 1395 mm 54 9 in Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation...
Page 135: ...able Floor Preparation Kits GE Healthcare supplied Illustration 5 32 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision...
Page 136: ...2285050 Refer to tables Section 2 3 Above grade Floor anchors without insert 2285051 Through bolts without insert 2285646 On grade Floor anchors 2286398 Above grade Floor anchors and insert 2285052 Th...
Page 137: ...ipping pallet to positioner dolly for table positioning used when the new baseplate is metric threads 2 3 2 2360538 Tilt Table Add On Kit Table 5 8 Item Name Part Description Quan Notes Omega Tilt com...
Page 138: ...ner and 4 anchors for Table 16 pc Above Grade anchor mount ing method hardware Grout assy 2285055 10 kg Powdered Mortar Ardex K 15 RTV Silicon Rubber Adhesive 18 mm making tape adhesive Open cell foam...
Page 139: ...ed Mortar Ardex K 15 RTV Silicon Rubber Adhesive 18 mm making tape adhesive Open cell foam 1 kit Used in constructing LC Po sitioner grout dam Dowel 2290937 2 Wood Dowel 16 mm diameter 12 pc Cable Con...
Page 140: ...thod hardware Washer Flat 99142204 Special Flat Washer for Through Bolts one for each bolt 16 pc same as above Plate 2290941 Special Steel Spacer Plate 4 in x 4 in 102 mm x 102 mm one for each bolt 16...
Page 141: ...n the floor plate is installed This part isn t GE part It s ordered locally 2 3 4 2285056 A C N Templates Table 5 15 Item Name Part Description Quan Notes A C N Tem plates 2185979 LC Positioner and Om...
Page 142: ...n 5 34 Baseplate plan 2359194 Angio configuration Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 142 2 Mounting R...
Page 143: ...Baseplate plan 2359194 Cardio configuration Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 5 Physical Ch...
Page 144: ...ever install injector and radiation protection on the same table accessory rail Typical installation on the front table accessory rail is Smart handle or Smart box Table Side System Control TSSC Innov...
Page 145: ...rals and respective power strip are not shown All have a single phase line a neutral line and a ground wire Illustration 6 1 Power distribution CE Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging...
Page 146: ...R rectifier 5 x 10 mm 3ph N PE AWG3 AWG6 3ph N PE 24 m 78ft 8in L out load 5 x 10 mm 3ph N PE AWG3 AWG6 3ph N PE 24 m 78ft 8in 1 Size and type may be adapted locally per local regulation Innova 2121 I...
Page 147: ...r supply cables from the other cables 2 1 Power Distribution Box CE Illustration 6 3 PDB CE 50 Hz Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Dir...
Page 148: ...lustration 6 4 PDB Schematic CE 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 148 2 Power Distribution Syste...
Page 149: ...stration 6 5 PDB Schematic CE 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 6 Electrical Connections...
Page 150: ...ution Box UL Illustration 6 6 PDB UL 60 Hz US only Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 150 2 Power Dis...
Page 151: ...stration 6 7 PDB Schematic UL 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 6 Electrical Connections...
Page 152: ...lustration 6 8 PDB Schematic UL 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 152 2 Power Distribution Syste...
Page 153: ...shing an equipotential grounding system refer to Grounding Continuity Job card in the Installation manual NOTICE All shielded cables shall have a connection to ground at each extremity without regard...
Page 154: ...Illustration 6 9 Grounding distribution Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 154 3 Grounding...
Page 155: ...801A C1 Frontal Cabinet Frontal Detector PWS AWG16 28802A C1 Lateral Cabinet Frontal Detector PWS AWG16 28803A C1 Frontal Cabinet Control Booth Power Strip 1 AWG6 28804A C1 Frontal Cabinet Control Boo...
Page 156: ...Chiller 6 mm AWG10 hospital PDB Ground 35 mm mini AWG2 0 mini hospital PDB Injector AWG10 hospital PDB Lateral XRT Chiller 6 mm AWG10 Table rail grounding bar 2268160 Table right rail Table left rail...
Page 157: ...er systems Cables shall be in conformity with local regulation UL CSA IEC CCC Table 6 2 Max Line Impedance for feeder line between Generator cabinet and Hospital V 380 400 420 440 460 480 Ohms 0 09 0...
Page 158: ...eder run length 10 m 35 35 35 35 35 35 35 35 35 20 m 35 35 70 35 35 50 35 35 50 30 m 50 70 95 35 50 70 35 50 70 40 m 50 70 120 50 70 95 50 70 95 50 m 70 95 185 70 95 120 70 95 120 60 m 95 120 240 70 9...
Page 159: ...ipment Room and Control Room Illustration 6 10 5 2 IVUS Floor Conduits A 3 inches and max 25 m long conduit between Control and Patient Rooms shall be prepared below the floor for the IVUS Innova inte...
Page 160: ...ial cable Volcano ID 806380001 1pcs Ground Cable 30 m 98 ft 5 in s5i Volcano ID 806889001 1pcs IVUS preinstallation Manual Volcano ID 806365003 Install 806889001 Ground Cable through the Control Booth...
Page 161: ...located in the control room suitable to the IVUS components positioning The maximum rated power of the Safety Isolation Transformer is 600 VA the maximum continuous system power consumption for s5i a...
Page 162: ...s to DICOM server The Volcano s5i supports the ultrasound multi frame image storage SOP class as an SCU service class user For more information see the Volcano s5 DICOM Conformance Statement located o...
Page 163: ...loor conduit prepared for the Control Booth cables between Equipment Room and Control Room Illustration 6 11 6 2 IVUS Floor Conduits A 3 inches and max 25 m long conduit between Control and Patient Ro...
Page 164: ...dures below Unpack and check the completeness of the Preinstallation kit 1 pcs Custom PIM cable 30 m 98 ft 5 in Volcano ID 806452001 1 pcs CAT5 cable 30 m 98 ft 5 in Volcano ID 807055001 1 pcs Custom...
Page 165: ...he connectors are properly protected during the rest of the installation 6 4 Electrical Requirements The power for the IVUS hardware components delivered with the product PC monitor Video Switch and p...
Page 166: ...to DVD The archived images are stored in DICOM format with Volcano s5i system acting as a File Set Creator FSC following the guidelines in the 2004 DICOM 3 0 specification Sending patient cases to DIC...
Page 167: ...ON LAMP MUST BE INSTALLED IN THE EXAM ROOM TO CONFORM TO INTERVENTIONAL STANDARD IEC EN 60601 2 43 SIGNAL INDICATING THE X RAY ON SHALL BE PERCEPTIBLE BY THE OPERATOR IN ALL THE LOCATIONS DEFINED FOR...
Page 168: ...cabinet Illustration 6 13 X Ray READY cabling between C2 cabinet and PDB NOTE Check local regulations if this indication is required If needed the lamp needs to be provided by the hospital or their c...
Page 169: ...provided by the hospital or contractor The wire size to connect to the C2 cabinet is 1 5 mm2 AWG 14 Relay rating is max 600 VAC Illustration 6 14 ROOM LIGHT cabling between C2 cabinet and PDB Innova 2...
Page 170: ...the FE for maintenance operations and to medical staff to adjust the tone signal level and to be able to see the power LED indicator Illustration 6 15 WARNING THE REMOTE SPEAKER KIT SHALL NOT BE INSTA...
Page 171: ...of conflict follow local regulation No other measures employed for radiation protection should cause the interruption of irradiation and any other disturbance of a procedure in progress Illustration 6...
Page 172: ...minimum of 150 kVA is required to supply the system in record mode Such an UPS would provide to the customer about 10 minutes of autonomy The UPS output shall follow the same output specifications as...
Page 173: ...onnect room device The injector must be connected to this supply 13 2 Pedestal Injector A power cord is supplied with the injector In all cases 110 VAC or 230 VAC it will be connected to a wall outlet...
Page 174: ...W AW spliter and monitors ground are equipotential 0 1 ohm equipotentiallity between room ground point busbar and wall outlets NOTICE To power the AW splitter use the appropiate cable out of lot P N 2...
Page 175: ...Electrical Ducts in Cable Channeling For MAC LAB System EX refer to Marquette document P N 2000465 001 Preinstallation Guide Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installat...
Page 176: ...nnova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Mis Charts General Guidelines Innova System introduces a new system interconnect with a star distribution for all cables from the technical...
Page 177: ...pass the cable and connector through with all other cables already in the conduit Also the size of conduit chosen must allow for future growth There is the possibility of additional cables being added...
Page 178: ...no gases box Floor mount installation can be accomplished one of two ways Connectors mounted in trough under mounting bracket Figure 1 Connectors mounted above mounting bracket Figure 2 Illustration...
Page 179: ...ration 6 18 Floor ducts and outlets Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 6 Electrical Connectio...
Page 180: ...aced on site For a description of how to use the following 5 cable group schematics see below Illustration 6 19 Description of cable group diagrams Cable length data is as follows Cable Length the tot...
Page 181: ...CABLE GROUP 1 FROM TECHNICAL AREA TO EXAM ROOM Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 6 Electrica...
Page 182: ...n 6 21 CABLE GROUP 2 FROM TECHNICAL AREA TO CONTROL ROOM Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 182 18 Ph...
Page 183: ...BLE GROUP 3 FROM TECHNICAL AREA TO TECHNICAL AREA Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 6 Electr...
Page 184: ...on 6 23 CABLE GROUP 4 FROM TECHNICAL AREA TO EXAM AREA Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 184 18 Phys...
Page 185: ...CABLE GROUP 5 FROM TECHNICAL AREA TO EXAM AREA Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 6 Electrica...
Page 186: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 186 18 Physical Runs...
Page 187: ...C2 Cabinet 1800 70 8 900 35 4 650 25 6 904 410 Omega Table Base Assembly 1240 49 960 38 2140 84 2 1 290 585 On pallet See Illustra tion 7 7 Omega Table Top Assembly 220 9 3470 137 840 33 155 70 On pa...
Page 188: ...ation 1 2 1 Innova LC Positioner Gantry On Shipping Dolly Illustration 7 1 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Re...
Page 189: ...Positioner Air Shipment Illustration 7 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 7 Additional Pla...
Page 190: ...ensions above are for shipping When the LP4 and dolly are in rolling configuration for hospital access the dimensions are Height 2001 mm 78 8 in and width 1040 mm 41 in Innova 2121 IQ 3131 IQ Biplane...
Page 191: ...ration 7 4 On shorter type dolly Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 7 Additional Planning Aid...
Page 192: ...Positioner Stationary Rails Packaging Illustration 7 5 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 192 1 Ship...
Page 193: ...bout 36 7 kg 81 lbs per dolly NOTE Please use appropriate transportation and lifting means when handling the system cabinets Professional handling is required Innova 2121 IQ 3131 IQ Biplane Cardiovasc...
Page 194: ...1 2 6 Omega Shipment Illustration 7 7 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 194 1 Shipping Information...
Page 195: ...ustration 7 8 1 2 8 Power Distribution Box PDB UL Illustration 7 9 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1...
Page 196: ...are shipped inside there original boxes group as needed on pallets Illustration 7 10 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 51779...
Page 197: ...Spanner 46 176584P1 High Voltage Cable In stallation Laptop Computer MS DOS Windows Positioner Configura tion and Calibration Innova LP4 Positioner Ladders Installation LP4 Toolcase shipped with LP4...
Page 198: ...livery of system DL User parts Monitor Suspension Ladders Installation XT Lifting Tool x2 46 156940G2 Installation Chiller Phillips Flathead screwdriver Open chiller Install wiring and hoses Innova 21...
Page 199: ...ata concerning the intended delivery route are recorded on the Route Survey in the following pages Record all loading capacities corridor widths door openings turning radii flooring materials elevator...
Page 200: ...y turn elevator ob struction etc Limitations loading capacity height width depth flooring material radius etc Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE He...
Page 201: ...tion Specialist on Site All cable and component selections must be sent at least 6 weeks before RSDD to your PMC The selection tool is available on the website http gein euro med ge com european_insta...
Page 202: ...to installation or storage area meets requirements and has been discussed and scheduled with the custom er Ensure floor protection is discussed requirements identified and will be avail able at time o...
Page 203: ...construction drawings or PMI speci fied installation drawings are required X X X X 10 Conduit electrical cable ducting dividers access flooring installed with the excep tion of surface mounted floor d...
Page 204: ...nnova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 204 5 Pre Installation Checklist...
Page 205: ...xt for obtaining an IP Address If you have questions or need clarification regarding the use of this form do not hesitate to ask the Operation support OnLine Contact OnLine Center Europe Telephone 33...
Page 206: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 206 1 IP Addressing Process...
Page 207: ...er from harware supplied with system Any additional hardware that is required will be the responsibility of the institution and or their contractor Contact your Installation Specialist with any relate...
Page 208: ...t UPS model 9130 Section 2 9 LC Positioner Section 2 10 LP4 Positioner Section 2 11 Power Distribution Box PDB CE Section 2 12 Power Distribution Box PDB UL Section 2 13 Omega IV Table Section 2 14 Om...
Page 209: ...stration 9 1 C1 frontal cabinet Slab on grade Upper floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter...
Page 210: ...net Illustration 9 2 C1 Lateral cabinet Slab on Grade Upper Floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1...
Page 211: ...stration 9 3 C2 LP LC cabinet Slab on floor Upper floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9...
Page 212: ...0 Chiller Illustration 9 4 Coolix 4000 chiller Slab on grade Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 212 2...
Page 213: ...ion 9 5 Coolix 4000 chiller Upper floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calculat...
Page 214: ...r Chiller Illustration 9 6 Thermocon Detector Chiller Slab on grade Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision...
Page 215: ...9 7 Thermocon Detector Chiller Upper floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calc...
Page 216: ...E Cabinet Illustration 9 8 Fluoro UPS CE cabinet Slab on grade Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 216...
Page 217: ...on 9 9 Fluoro UPS CE cabinet Upper floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calcula...
Page 218: ...L Cabinet Illustration 9 10 Fluoro UPS UL cabinet Slab on grade Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 21...
Page 219: ...on 9 11 Fluoro UPS UL cabinet Upper floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calcul...
Page 220: ...odel 9125 Illustration 9 12 3 kVA Cabinet UPS model 9125 Slab on grade Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revisi...
Page 221: ...13 3 kVA Cabinet UPS model 9125 Upper Floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Cal...
Page 222: ...odel 9130 Illustration 9 14 3 kVA Cabinet UPS model 9130 Slab on grade Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revisi...
Page 223: ...15 3 kVA Cabinet UPS model 9130 Elevated Floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic C...
Page 224: ...itioner Illustration 9 16 LC Positioner Slab on grade 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 224 2 Ca...
Page 225: ...ion 9 17 LC Positioner Slab on grade 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calcula...
Page 226: ...Illustration 9 18 LC Positioner Upper floor 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 226 2 Calculations...
Page 227: ...tion 9 19 LC Positioner Upper floor 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calculat...
Page 228: ...LP4 Positioner Illustration 9 20 LP4 Positioner 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 228 2 Calculat...
Page 229: ...ustration 9 21 LP4 Positioner 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calculations 2...
Page 230: ...el PDB CE Illustration 9 22 Main Disconnect Panel PDB CE Slab on grade Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revisi...
Page 231: ...23 Main Disconnect Panel PDB CE Upper floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Cal...
Page 232: ...el PDB UL Illustration 9 24 Main Disconnect Panel PDB UL Slab on grade Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revisi...
Page 233: ...25 Main Disconnect Panel PDB UL Upper floor Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Cal...
Page 234: ...V Table Illustration 9 26 Omega IV Table Slab on grade 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 234 2 C...
Page 235: ...on 9 27 Omega IV Table Slab on grade 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calcula...
Page 236: ...llustration 9 28 Omega IV Table Upper floor 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 236 2 Calculations...
Page 237: ...ion 9 29 Omega IV Table Upper floor 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calculat...
Page 238: ...Baseplate Illustration 9 30 Omega V Long table w baseplate Slab on grade 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN...
Page 239: ...Omega V Long table w baseplate Slab on grade 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic...
Page 240: ...ion 9 32 Omega V Long table w baseplate Upper floor 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 240 2 Calc...
Page 241: ...Omega V Long table w baseplate Upper floor 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic...
Page 242: ...g Table Illustration 9 34 Omega V long table Slab on grade 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 242...
Page 243: ...9 35 Omega V Long table Slab on grade 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calcu...
Page 244: ...ustration 9 36 Omega V Long table Upper floor 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 244 2 Calculatio...
Page 245: ...n 9 37 Omega V Long table Upper floor 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calcul...
Page 246: ...ension Ceiling mounted Illustration 9 38 Overhead Flat Panel Suspension 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN R...
Page 247: ...n 9 39 Overhead Flat panel suspension 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seismic Calcul...
Page 248: ...eiling Mounted Illustration 9 40 8 Monitor Flat panel Suspension ceiling mounted 1 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 517795...
Page 249: ...Monitor Flat panel Suspension ceiling mounted 2 2 Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 Chapter 9 Seism...
Page 250: ...Innova 2121 IQ 3131 IQ Biplane Cardiovascular Imaging System Pre Installation Manual GE Healthcare Direction 5177951 4 1EN Revision 1 This page left intentionally blank 250 2 Calculations...
Page 251: ......
Page 252: ...2010 General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare 283 rue de la Mini re 78530 Buc FRANCE www gehealthcare com...