NOTICE
The Substructure for Dual Arm suspension is mandatory to install the MAVIG
suspension with fixed point dual arm.
NOTICE
The lower edge of the Substructure for Dual Arm suspension should be the same
height as the lower edge of the false ceiling.
Substructure for Dual Arm suspension
The Substructure for Dual Arm suspension is used to attach the MAVIG suspension with fixed point dual
arm to the solid ceiling. It is used as the bridging element between the solid ceiling and the false ceiling
for the installation and the fixation of the suspension.
Also, it provides a hooking point required for the installation and the replacement of the Large Display
Monitor.
The Substructure for Dual Arm suspension is mandatory to install the MAVIG suspension with fixed point
dual arm for Non-seismic Zones. For Seismic Zone installations, refer to Structural Engineer for
appropriate design of the structure for installing the MAVIG suspension system.
For standard site configurations, the distance between the ceiling and the lower edge of the false ceiling
should be in a range of minimum 175 mm and maximum 610 mm.
If the distance between the ceiling and the false ceiling is less than 175 mm, then the middle plate is not
installed.
• Distance between ceiling and false ceiling: Minimum is 175 mm and maximum is 610 mm
• Distance between ceiling and false ceiling: Less than 175 mm
Item
Description
[1]
Weight in kg
[2]
Ceiling Plate
Equipment Requirements
112
Innova
TM
IGS 6
5750182-1EN 3
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...