[1] : Port diameter of the false ceiling: maximum is 620 mm.
2.3.4 Wall Requirements
The optional LD secondary monitor outside the patient room is mounted on wall. The swingout arm that
holds the LD additional monitor shall be mount according to the manufacturer mounting manual, see
Articulating Arm Wall Mount Installation Manual in OEM manuals list.
An hooking point shall be provided in order to uplift the LD additional monitor at the swingout arm level
during its installation:
• Hooking point characteristic: It must withstand up to 440 lbs (200 Kgs)
• Hooking point position:
Figure 2-65 Hooking point position
• Recommended hooking point dimensions:
Figure 2-66 Hooking point dimensions
The position of the Swing out arm Center of Gravity (with weights) is given below.
Equipment Requirements
5750182-1EN 3
Innova
TM
IGS 6
115
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...