Seismic Calculations
Seismic requirements are determined and specified by the hospital/ Design Professional of record and
may require approval by the specific state or country agency.
Seismic attachment hardware shown on seismic calculations may differ from hardware supplied with
system. Any additional hardware that is required will be the responsibility of the institution and/or their
contractor. Contact your local GE Installation Program Manager to obtain seismic calculations.
Seismic calculations are per California Building Code (CBC) and International Building Code (IBC).
Large Display Monitor suspension with fixed point dual arm
CAUTION
THE STANDARD SUBSTRUCTURE (MAVIG GD60D022) SHOULD NOT BE USED WITH
SYSTEM IN SEISMIC ZONE.
Contact MAVIG or local contractor to design and supply specific substructure
including M12 threaded holes requirement (see below).
Four M12 threaded holes with hooking point are required for the installation of the dual arm suspension,
the installation and replacement of the Large Display Monitor. The structural support plate (
Center of gravity and seismic support, and attachments on page 118
) should include these 4 x M12
threaded holes.
For the threaded holes positioning on the structural support plate refer to
Substructure for Dual Arm suspension dimensions on page 57
in Dimension Drawings.
Equipment Requirements
5750182-1EN 3
Innova
TM
IGS 6
117
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...