NOTE
(1)
: The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are
6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66
MHz to 40,70 MHz
(2)
: Field strengths from fixed transmitters, such as base stations for cellular telephones
and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast
cannot be theoretically estimated accurately. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be performed. If the
measured field strength exceeds the RF compliance level above, observe the Innova
system to verify normal operation in each use location. If abnormal performance is
observed, additional measures may be necessary, such as re–orienting or relocating the
Innova system.
(3)
: Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
(4)
: Refer to table & warning below:
WARNING
THE INNOVA IGS SYSTEM IS A LARGE, PERMANENTLY-INSTALLED MEDICAL
EQUIPMENT FOR WHICH THE SIMULATED OPERATION IN AN ANECHOIC CHAMBER
IS NOT FEASIBLE AND CONSEQUENTLY IS EXEMPT FROM THE TESTING
REQUIREMENT SPECIFIED BY IEC 61000-4-3.
THE INNOVA SYSTEM HAS NOT BEEN TESTED FOR RADIATED RF IMMUNITY OVER
THE ENTIRE FREQUENCY RANGE 80 MHZ TO 6 GHZ.
THE INNOVA SYSTEM HAS BEEN TESTED FOR RADIATED RF IMMUNITY ONLY AT
SELECTED FREQUENCIES, AND USE NEARBY OF EMITTERS AT OTHER
FREQUENCIES COULD RESULT IN IMPROPER OPERATION.
Table 3-4 IEC60601-1-2 ed2.1 &3 field level & frequencies
Tested Frequencies (MHz)
Field Level (V/m)
Modulation
433.920 (ISM)
(1)
3
80% AM at 1 kHz rate
915 (ISM)
(1)
1440
1750
1920
2450 (ISM)
(1)
NOTE
(1)
: Industrial, Scientific and Medical (ISM) radio bands.
The associated recommended separation distances as per IEC60601-1-2 ed2.1 & 3 are listed in
Special Construction Requirements
5750182-1EN 3
Innova
TM
IGS 6
123
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...