indicates the typical heat dissipation at nominal load and specify cooling air flow
in two situations:
• 1000 m (3280 ft) altitude, cooling air between 25°C (77°F) and 30°C(86°F)
• 250 m (820 ft) altitude, cooling air below 18°C (64°F)
Table 4-2
Fluoro UPS
Typical heat output
On-line cooling air flow
at 30°C / 86°F
Typical on-line cooling
air flow at 18°C / 64°F
CE version
1.50 kW (5101 BTU/hr)
625 m
3
/h (368 CFM)
227 m
3
/h (134 CFM)
UL version
4720 BTU/hr (1.38 kW)
301 CFM (511 m
3
/h)
119 CFM (203 m
3
/h)
4.4 Acoustic Specifications
Audible noise:
• Less than 50 dB (A) at 1 meter for an Innova Frontal Positioner.
• Less than 50 dB (A) at 1 meter for an Innova Lateral Positioner.
• Limited to 50 dB (A) at 1 meter for Omega V table.
• Limited to 55 dB (A) at 1 meter for C2 Cabinet (Frontal/Lateral).
• Limited to 60 dB (A) at 1 meter for the Coolix 4100 chiller.
• Limited to 65 dB (A) at 1 meter for C1 Frontal Cabinet.
• Limited to 65 dB (A) at 1 meter for C1 Lateral Cabinet.
• Limited to 52 dB (A) (background of 35 dB (A)) at 1 meter for Digital Detector Conditioner Thermo–Con.
• Less than 50 dB (A) at 1 meter for a DL LCD monitor.
• Less than 50 dB (A) at 1 meter for the Fluoro UPS.
The values contained in
Table 4-1 Gantry Noise measurements on page 134
are the results of the
measurement performed on the first system with 31 cm Revolution Digital detector installed on a clinical
site.
Table 4-1 Gantry Noise measurements
Measurements
Gantry (FRT, LAT, both)
Gantry posi-
tions
Peak (db)
Mean (db)
Ambient (System off)
1
N/A
55.7
System ON (no motion, no
X-Ray)
2
N/A
58.2
Motions
FRT (Detector lift)
1
86.2
66.3
LAT (C-Arc)
2
86.3
63.7
FRT and LAT
2
85.1
66.9
Environmental Requirements
134
Innova
TM
IGS 6
5750182-1EN 3
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...