NOTICE
Battery safety instructions:
• The battery–voltage is dangerous for person’s safety.
• Never dispose of battery in a fire: They may explode.
• Do not open or mutilate battery: Their contents (electrolyte) may be extremely
toxic. If exposed to electrolyte, wash immediately with plenty of water.
• Avoid charging in a sealed container.
Full UPS Record (compatibility)
An UPS sized for a minimum of 160 kVA is required to supply the system in record mode. Such an UPS
would provide to the customer about 10 minutes of autonomy.
The UPS output shall follow the same output specifications as shown in Optional Isolation transformer
specifications in
5.1.1 Electrical Requirements on page 137
.
Contact your local GE Healthcare representative for more information and/or public power distribution
analysis.
System failure
In the event of a system failure with a patient on table during an examination, the operator can utilize a
Surgical Imaging mobile unit to finish the examination.
In this case a wall outlet single phase + ground is required to feed the mobile. It is also requires a free
space around the patient table to proceed with the mobile instead of Frontal Positioner. The table has to
rotate to 90º. The minimum room width of 4400 mm (14.5 in) may not be sufficient for this scenario.
5.1.4 Power Distribution System
NOTICE
It is recommended to separate power supply cables from the other cables.
Electrical Requirements
5750182-1EN 3
Innova
TM
IGS 6
145
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...