Figure 5-1
WARNING
THE REMOTE SPEAKER KIT SHALL NOT BE INSTALLED IN THE PATIENT VICINITY.
Patient Vicinity Definition : see Patient Environment Equipment in
2.2.3 Room Layout Considerations on
5.9 MAC-LAB System EX
For location of MAC–LAB Acquisition Unit Floor mount kit 408431–001, refer to Electrical Ducts in
5.3.2 Cable Channeling on page 156
For MAC–LAB System EX, refer to:
- Marquette document P/N 2000465–001 Preinstallation Guide
5.10 Advantage Windows
A power cord is supplied with Advantage Windows. It is connected to outlet fed by the main disconnect
room device.
Verify that the ground between AW, AW spliter and monitors ground are equipotential (0.1 ohm
equipotentiallity between room ground point (busbar) and wall outlets).
Electrical Requirements
176
Innova
TM
IGS 6
5750182-1EN 3