Fluoro UPS
The Fluoro UPS solution allows the customer to complete an exam in fluoroscopy mode in case of mains
power failure. The autonomy provided by this UPS is 5 minutes of fluoroscopy every 24 hours. It includes
the following components:
• Fluoro UPS (2 different model for UL and CE markets),
• EMI filter box (CE market only).
Monitor suspensions
Six monitors suspension
The common type of this suspension is an XT inboard monitor bridge. A monitor frame receives 6
monitors: 4 monitors for Innova images and 2 color monitors for 3rd party images or optional Innova
Images (simultaneous display option).
This suspension is delivered and installed by GE.
LDM suspension
For the systems with the LDM option, a specific suspension can be provided:
• suspension with an XT inboard monitor bridge
• suspension with fixed point dual arm
These suspensions are delivered and installed by GE.
The two backup monitors are mounted on the back of this suspension for faster access in case of failure.
For the second LDM, a wall mounting kit can be provided.
Third party monitor suspension
In addition or replacement of the GE provided suspensions, the customer can install the suspension of his
choice (“third party monitor suspension”), provided the requirements below are met:
• The suspension shall be installed by strictly following the GEHC installation instructions. GE Healthcare
specifically disclaims any and all liability arising out of or relating to the use or performance of the
monitor suspension (including cables), including, without limitation, any liability or claims relating to
patient injury, death, or the reliability of such monitors suspension(s).
• The suspension(s) shall comply with the applicable Regulation enforced in the country (eg., when
installed in an European Community country, it shall be CE marked).
• All Regulatory Requirements and Pre-Installation Instructions contained in Third Party Monitor
Suspensions – Service Instruction for Installation contained in the Service manual shall be met.
The association of the Innova product and the purchaser’s (customer) monitors suspension(s), is not
covered by the product certification.
Equipment Requirements
5750182-1EN 3
Innova
TM
IGS 6
35
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...