System Compatibilities Cross-Reference – 19” LCD Monitors Sub-System
Product Name
Pre-Installation Document Num-
ber
Catalog Number
LCD 19” EIZO MX193 Color Monitor
provided by EIZO
Eizo 19” LCD HB color monitor
RX150 GE
sm 5499528-1-8EN
2.2 Room Layouts
2.2.1 Room Dimension Requirements
For Room size dimensions, refer to
2.2.2.1 Patient Room Layout on page 72
.
For additional details, refer to
2.2.3 Room Layout Considerations on page 80
.
2.2.2 Room Layout Drawings
2.2.2.1 Patient Room Layout
WARNING
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS MANDATORY. THE
ELECTRONIC CABINETS (C1, C2, OPTIONAL LD CABINET, COOLIX 4100 CHILLER,
DETECTOR CONDITIONER AND FLUORO UPS WHEN INSTALLED ) INCLUDE FANS
THAT ARE CREATING AIR-CIRCULATION OF PULSED-AIR. WHEN THIS PULSED AIR
IS IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN
EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF TRANSMISSION OF THESE
AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL
PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED IN A
ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E., TECHNICAL ROOM.
Table 2-8 Exam room dimensions
Room Dimensions
Length x Width
Ceiling Height
Recommended:
11570 mm x 8000 mm (38 ft 0 in x 26 ft 2
in)
2845 mm ± 5 mm (9 ft 4 in ± 0.2 in) is man-
datory ceiling height
Minimum:
6900 mm x 4400 mm (22 ft 8 in x 14 ft 5 in)
Equipment Requirements
72
Innova
TM
IGS 6
5750182-1EN 3
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...