Table 3-2
Immunity
Test
IEC 60601–1–
2 Ed2.1 & 3
Test Level
IEC 60601–1–
2 Ed4 Test
Level (profes-
sional
healthcare
environment)
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge
(ESD)
IEC 61000-4-2
+/-6 kV con-
tact
+/-8 kV air
+/-6 kV con-
tact
+/-8 kV air
+/-6 kV con-
tact
+/-8 kV air
Floors are wood, concrete, or ceramic tile, or
floors are covered with synthetic material and
the relative humidity is at least 30 %.
Electrical fast
transient/
burst
IEC 61000-4-4
+/-2 kV for
power supply
lines
+/-1 kV for in-
put/output
lines
5 kHz burst
repetition fre-
quency
+/-2 kV for
power supply
lines
+/-1 kV for in-
put/output
lines
100 kHz burst
repetition fre-
quency
+/-2 kV for
power supply
lines
+/-1 kV for in-
put/output
lines
5 kHz & 100
kHz burst rep-
etition fre-
quency
Mains power quality is that of a typical com-
mercial or hospital environment.
Surge
IEC 61000-4-5
+/-1 kV line(s)
to lines(s)
+/-2 kV line(s)
to earth
+/-1 kV line(s)
to lines(s)
+/-2 kV line(s)
to earth
+/-1 kV line(s)
to lines(s)
+/-2 kV line(s)
to earth
Mains power quality is that of a typical com-
mercial or hospital environment.
Voltage dips,
short inter-
ruptions and
voltage varia-
tions on pow-
er supply in-
put lines
IEC
61000-4-11
< 5% U
T
(>95% dip in
U
T
) 0% for
5sec
0 % U
T
250/300 cycle
< 5% U
T
(>95% dip in
U
T
) 0% for
5sec
0% U
T
250/300 cycle
Mains power quality is that of a typical com-
mercial or hospital environment. If the user of
the Innova system requires continued opera-
tion during power mains interruptions, it is
recommended that the Innova system be
powered from an uninterruptible power sup-
ply.
Power fre-
quency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields are at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note: U
T
is the mains voltage prior to application of the test level. 250/300 cycle means 250 periods at 50Hz or
300 periods at 60Hz
Special Construction Requirements
5750182-1EN 3
Innova
TM
IGS 6
121
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...