The Innova system is intended for use in the electromagnetic environment specified below.
The purchaser or user of the Innova system should assure that it is used in such an environment.
Table 3-3
Immunity
Test
IEC 60601–1–2
Ed2.1 & 3 Test
Level
IEC 60601–1–2
Ed4 Test Level
(professional
healthcare envi-
ronment)
Compliance
Level
Electromagnetic Environment
Conducted
Radio Fre-
quency
IEC
61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms 150 kHz
to 80 MHz
6Vrms in ISM
bands
(1)
V
1
= 3 V
(1)
Portable and mobile RF communications
equipment is used no closer to any part
of the system, including cables, than the
recommended separation distance calcu-
lated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d = [3.5/V1]√P
d = [3.5/E1]√P, from 80 MHz to 800 MHz
d = [7/E1]√P, from 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
(2)
, are less than the compliance
level in each frequency range
(3)
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated Ra-
dio Frequen-
cy
IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
80 MHz to
2.7 GHz
E
1
= 3 V/m
(4)
Special Construction Requirements
122
Innova
TM
IGS 6
5750182-1EN 3
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...