Figure 5-1 X-Ray READY cabling between C2 cabinet and PDB
NOTE
Check local regulations if this indication is required.
If needed, the lamp needs to be provided by the hospital or their contractor.
5.7 Door Interlocks
CAUTION
IEC 60601–2–43 requires not to install door interlocks. It is the responsibility of
the field service to check that this requirement is not in contradiction with local
regulation. In case of conflict, follow local regulation. No other measures
employed for radiation protection should cause the interruption of irradiation
and any other disturbance of a procedure in progress.
Electrical Requirements