2--3--8 System Clock Adjust
Procedure
Follow these steps to adjust the system clock of the LOGIQ 200 PRO Series.
1.
Using the
Arrow
up/down keys, select the “System Clock Adjust” menu in the Top Menu. Refer to the
ILLUSTRATION 2--5.
2.
Press
Set
or
Return
key.
The System Clock Adjust Menu will appear on the monitor as shown in the ILLUSTRATION 2--28.
Service Utility MENU :
Top Menu
99/03/12 09:16:37
Current Software Version is 3.0
System Diagnosis
Board/Program Version
System Reset
Key Operation History
Option Set
Initialize :1
(1:America 2:Europe 3:Asia 4:Japan)
*System Clock Adjust
Software/Probe data upgrade
Exit
SYSTEM CLOCK ADJUST MENU
ILLUSTRATION 2--28
3.
Input current date and time, and press
Set
or
Return
key.