4-1
GE MEDICAL SYSTEMS
ULTRASOUND QUALITY ASSURANCE REFERENCE MANUAL
REV 1
2262684-100
SECTION 4
PROBE QA CHECKS
CAUTION
Possible risk of infection. Do not handle soiled or contaminated probes and other
components that have been in patient contact. Follow appropriate cleaning and
disinfection procedures described in the Introduction section of this manual, before
handling the equipment.
4-1
Probe Electrical Safety Measurements and Checks
Probe electrical safety tests must be the last procedure to be performed following re-assembly and rework.
Should any test fail to be within published limits or if results are noticed to be non typical, it is required to
troubleshoot the cause. Following identification and correction of the fault or cause, a complete retest of
the system for electrical safety is required.
•
Use the Electrical Safety Data table on the QA Procedure to record the values for the probes.
•
ANY PROBE FAILING ELECTRICAL LEAKAGE TESTS MUST BE REMOVED FROM SERVICE.
•
All Electrical safety items must be completed and within acceptable limits prior to sign-off by the
customer.
NOTE:
Disconnect any video cables external to the system that may provide an additional
ground Path.