GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
7 - 8
Section 7-4 - Screen Captures
5.) A Save dialog box will be opened. Choose
d:\export folder
as the archive location to save the
image on the hard disk, MOD or CD-R.
7-4-4
Reset the P4 Key to Customer’s Functionality
If the customer had programmed the P4 Key to a function other than screen capture, restore that
functionality recorded in section 7-4-1 on page 5. Refer to Figure 7-2.
1.) Select
Utilit
y from the Front Panel.
2.) Select
Connectivity
from the Menu.
3.) Select the
Buttons
tab on the Connectivity screen.
4.) In the
Button
field, select Print4.
5.) In the
Destinations
section, select the service(s) recorded in
. The
destinations list displays the following information:
*
Name: user defined during service configuration
*
Type: the type of service
*
Server: the device for which the service was configured
*
Dir: direction: output, input, or both (I+O)
NOTE:
Only output services can be associated to the print keys.
6.) In the
Image generated
section, select the parameters related to the service recorded in
, Section
Figure 7-5 Save Dialog Box
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...