Chapter 8 Replacement Procedures
8 - 29
8-4-7
TGC Knob Set (FRU No. 206)
Purpose: This is a description on how to remove and replace TGC Knobs.
8-4-7-1
Tools
•
Common pillips screwdrivers.
8-4-7-2
Needed Manpower
•
1persons, 5 m travel
8-4-7-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-4-7-4
Removal Procedure
1.) Remove eight (8) TGC knobs from the upper side of the Keyboard. Refer to Figure 8-29.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-4-7-5
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-29 Removing TGC Knobs
Table 8-14 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
2380207
, Rev 7+, Section 8-4-7. Equipment
passed all required tests and is ready for use. “
Section 4-11-4