Chapter 8 Replacement Procedures
8 - 35
8-4-10
OP Panel HUB board Assy (FRU No. 209)
Purpose: This is a description on how to remove and replace HUB board Assy.
8-4-10-1
Tools
•
Common pillips screwdrivers.
8-4-10-2
Needed Manpower
•
1persons, 5 m travel
8-4-10-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
8-4-10-4
Removal Procedure
1.) Unscrew five (5) screws and remove the EMC guard from the HUB Board. Refer to Figure 8-36.
2.) Disconnect six (6) connectors (1-6) from the HUB Board Assy. Refer to Figure 8-37.
Figure 8-36 Open EMC guard
Figure 8-37 Disconnect 6 connectors
1
2
3
4
5
1
2
3
4
5
6