7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 198
Section 8-11 - Installing Base Software Load Image
8-11-3-3
Functional Checks for Service Platform (Diagnostic Program)
NOTE:
After performing diagnostics, shut down and restart to make the system working properly.
1.) Make sure that Service button is shown at the top of the Utility screen. Click on the service button
to activate. It will take about ten (10) seconds for activating.
2.) Make sure that
CAPS
is not selected (should be dimmed) for password entry performed later.
NOTE:
If the following dialog box is shown on the monitor, select "Do not perform this check in the
future" and click on the Yes button. This dialog box will not be displayed for next time.
Figure 8-239 CAPS not selected
Figure 8-240 Netscape Navigator