GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 201
8-11-3-4
Setting the VCR (if necessary)
1.) Press
Utility > System>Peripheral.
2.) Set the proper parameters:
-
PAL or NTSC
-
VCR type
-
Source type
3.) Click on
Save.
4.) Shut down the scanner, then restart the scanner.
5.) Verify that the VCR can be properly operated (remote, play...).
Figure 8-244 Utility > System>Peripheral
NOTICE
Be sure to connect the converter and cable before booting up the system. Otherwise the LOGIQ 5
system does not recognize the peripheral devices
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...