4 - 52
Section 4-13 - Board Function Checks
4-13-11
BEP Assy function check procedure
4-13-11-1
HDD, Memory, CPU Diagnositc Test
1.) The Start up Screen will be shown on the Monitor display when the system is Turned ON.
2.) Figure 4-69 displays when Service Dongle inserted in Rear Panel.
3.) To enter the Maintenance Mode, select Maintenance button.
4.) To start Diagonistic Test, Execute PC Doctor.
'Start' => 'Programs' => 'PC Doctor'
5.) HDD Test
Select Hard driver test manu, start to testing(Figure 4-70).
Check display 'Pass' massage
Figure 4-69 Start Application window
Figure 4-70 HDD Test