Chapter 8 Replacement Procedures
8 - 27
8-4-6
OP Panel TGC Assy (FRU No. 205)
Purpose: This is a description on how to remove and replace OP Panel TGC Assy.
8-4-6-1
Tools
•
Common pillips screwdrivers.
8-4-6-2
Needed Manpower
•
1persons, 5 m travel
8-4-6-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
8-4-6-4
Removal Procedure
1.) Remove eight (8) TGC knobs from the upper side of the Keyboard. Refer to Figure 8-27.
2.) Disconnect the TGC connector from the back side of the keyboard.
3.) Unscrew four (4) screws and Remove the TGC Assy from the keyboard.
Figure 8-27 Removing TGC Knobs
Figure 8-28 Unscrew 4 screws and remove TGC Assy
TGC Connector
1
2
3
4