GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 91
8-7-2
LMT Assy (FRU No. 501)
Purpose: This is a description on how to remove and replace the LMT Assy.
8-7-2-1
Tools
•
Common pillips screwdrivers
8-7-2-2
Needed Manpower
•
1 person, 25 m travel
8-7-2-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-7-2-4
Removal Procedure
1.) Remove the RLY Assy. Refer to section 8-7-1 on page 88.
2.) Unscrew two (2) screws (1-2) to remove the LMT Board. Refer to Figure 8-82.
CAUTION
An electronic discharge may damage a component. Turn OFF power and wear the wrist strap
before you remove circuit boards. Do not unplug the power cord to keep ground continuity.
Do not bend or flex the boards when mounting/dismounting each boards. Surface mount IC
boards are very susceptible to damage from flex/torque.
Figure 8-82 Unscrew 2 screws to remove the LMT B’d
2
1
2
1
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...