Chapter 8 Replacement Procedures
8 - 157
4.) Unscrew two (2) screws (12-13) and disconnect twenty (20) connectors from the front side of the
LV unit using long-nose gripper. Refer to Figure 8-175.
5.) Remove the LV unit.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-9-2-5
Mounting Procedure
Install the new parts in the reverse order of removal.
8-9-2-6
DC Voltage level adjustment
1.) After LV unit replacement , the DC output level should be set proper level. Refer power supply
adjustment ,section Section 6-3 on page 1
Figure 8-175 Unscrew 4 screws and disconnect 20 connectors
Table 8-50 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
2380207
, Rev 7+, Section 8-9-2. Equipment
passed all required tests and is ready for use. “
LV Unit function check procedure
RIGHT SIDE VIEW
(12)
(13)