GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 192
Section 8-11 - Installing Base Software Load Image
8-11-1-3
DMC Checks (DMC User Only)
1.) Check DMC Cable (USB to Serial Bridge).
NOTE:
DMC Cable is supplied only as an option.
2.) Plug in the Cable as shown on the right side of the USB port. Refer to Figure 8-230.
3.) Ensure USB to Serial Bridge port is COM4.
4.) Trackball to My computer icon and left click
5.) Right click on Properties.
6.) Right click on the hardware tab.
7.) Right click on the device manager in the middle right of the hardware tab window.
8.) Expand ports (COM & LPT), If not COM4, Proceed as followed.
9.) .Double click “USB to Serial Bridge”.
Figure 8-230 DMC Cable
Figure 8-231 Device Manaer Window
DMC Cable
Plug on the right
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...