Advanced Cardiac Calculations
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
4–64
New Measurement Sequence
ORDER Command
After selecting the desired Main Menu and Sub-Menu, the
calculations are displayed in the top half of the menu.
Use the Trackball to highlight the desired calculation and press
Set. Measurement and calculation items are displayed on the
bottom half of the menu. The order in which they are executed
is shown to their left.
To change the order or omit undesired items from the
sequence, use the ORDER command. Trackball the highlight
cursor to ORDER and press Set.
All measurements and calculations on the bottom half of the
menu no longer have a number to their left.
Use the Trackball to move the highlight cursor to the first
measurement to be made in the customized sequence. Press
Set. The number one is placed to the left of that item. Use the
Trackball to move the highlighted cursor to the second item in
the new sequence and press Set.
Continue the process until all desired items are designated.
NOTE: Items not designated will be deleted from the sequence.
Use the SAVE command to store the new custom sequence.
S
The customized sequence is saved until the ORDER
command is used to generate a new sequence.
S
Items not designated will not automatically appear as a
measurement prompt.
RESET Command
The RESET command resets the measurement sequence to
the factory defaults. Prior to resetting, the following message is
displayed:
“Reset? Press ‘ESC’ to cancel. ‘SET’ to confirm”
Set resets the order to factory defaults. Esc causes nothing to
happen.
.
Hints